TECHNICAL SPECIFICATIONS:-
Sr. No. |
Main component |
Detailed Specifications |
1.0 | Gas Chromatography | GC with FID, NPD and ECD Detector along with original licensed windows-based software and Split-Spitless capillary inlet along with cool on column Inlet with Auto injector. GC must be capable to accommodate at least four two detectors & two injectors in working conditions simultaneously. System should support retention time locking facility with EPC resolution 0f 0.001 psi. |
1.1 | Column Oven | ·Capable of housing at least two columns; operating temperature range: Upto 450 ° or more.
·Temperature Programming Ramps: 20 or more. ·Temperature set point resolution 0.1-degree C or more. |
1.2 | Pneumatics | ·System must have pneumatic Electronic Flow Control for all injectors & detectors. |
1.3 | Inlets (Two) | · Split/split less capillary port
·Temperature: 400 °C or more ·Fully EPC · Split ratio: 6000: 1 or more · Pressure setting range 0–100 psi · Cool on column Inlet ·Temperature 400 °C or more ·Pressure setting range 0–100 psi |
1.4 | Nitrogen Phosphorus Detector (NPD) | ·Minimum detectable level: 0.1 pg N/s and 0.05 pf P/s
·Linear dynamic range: >105 ·Maximum temperature: 400 °C ·Data acquisition rate up to 300 Hz |
1.5 | Flame Ionization detector (FID)
 |
·Electronic pressure and flow control of all gases
·Maximum temperature: 400 °C or more ·Minimum Detectivity: 1.5 pg C/sec or less ·Data acquisition rate up to 300 Hz ·Linear dynamic range: 107 |
1.6 | Electron Capture Detector (ECD) | ·Electronic pressure and flow control of all gases
·Maximum temperature: 400 °C or more ·Minimum Detectivity: 3.8 fg/ml or less ·Data acquisition rate up to 50 Hz ·Linear dynamic range: 104 |
1.7 | Auto Sampler | ·It should be able to do liquid injection with 2 ml vials.
· 50 or more Vials capacity. |
1.8 | Software |
· Software should offer extensive custom calculation capabilities (Concentration, calibration curve, peak soothing, peak subtraction etc.) eliminating the need for external applications to meet the laboratory need. |
2.0 | Capillary Columns | · Column for pesticides analysis. – Qty 2
· 100-meter column for Fatty acid. – Qty 2 |
3.0 | Consumables | ·Vials and Caps- 500
·Liner- 10 each for split and split-less ·Ferrules- 10 ·Column nut- 10 ·Septa- 100 · Auto sampler syringe- 4 |
4.0 | Computer | ·Compatible PC with 8 GB RAM, 512 GB HDD and monitor. |
4.1 | Printer | · Laser Jet Printer |
4.2 | Standards | ·FAME standards |
5.0 | Gas Cylinders | ·Filled Gas Cylinders of Hydrogen, Nitrogen and Zero Air with Gas Purification Panel and double stage brass regulators. |
5.1 | Delivery | ·Delivery till F.O.R Destination |
5.2 | Instrument performance | ·Installation at our site with detailed demonstration of all performance specifications including repeatability, software features, do’s and don’ts etc. |
6 | IQ/ OQ | ·IQ, OQ, PQ of instrument and Software should be provided along with document. |
7 | On-site training | ·Site Training for 2-3 working days on operation and troubleshooting should be part of the installation and commissioning of the system |
8 | Warranty | ·One year warranty from the date of installation.
·The vendor should have availability for ten years guaranteed parts and services. ·Current user’s / performance list in India of the quoted Model with contact details (Customer name, phone email id etc) and date of installation to be provided (Minimum 5 installations of the model quoted)and this will be taken very seriously while making the decision. Your post sales service feedback will be certainly a deciding factor.
|
9 | Compliance Statement |
·        All pre-installation requirements must be attached |